• CDs

    • What detergents and their respective concentrations are compatible with Gyrolab Bioaffy CDs?
    • Some detergents may affect the function of the hydrophobic barrier. High concentrations of detergents should be avoided since the volume definition function may not function properly. This might affect the precision of the assay. The chemical compatibility for all Gyrolab Bioaffy CDs is as follows: 0.2% Tween-20, 40% methanol, 30% dimethyl sulfoxide (DMSO), 30% 2-propanol, 0.01% Triton X-100, 0.1% SDS.

    • What is the size of the affinity-capture column?
    • 15 nl.

    • What are the storage conditions for Gyrolab Bioaffy CDs and Gyrolab Mixing CD?
    • An unopened package should be stored at +4°C to +8°C. A partially used CD saved for additional analysis of unused microstructures can be stored dark, dry and at room temperature for one week, in the re-sealed pouch and in the CD box.